Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Yokogawa uR100 Zero and Span CAL ADJ /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 70 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD ADJUSTMENT LOCATION 1.002 DISP Leave internal assembly pulled out. The adjustments 1.002 DISP are located on the left side of the unit. 1.002 DISP 1.002 DISP Channel 1 adjustments are the TOP set, channel 2 are 1.002 DISP under that, and channel 3 are under those. 1.003 HEAD CHANNEL [MEM] ZERO and SPAN Adjustments 1.004 DISP Using the keyboard on the side of the UUT... 1.004 DISP 1.004 DISP Set the CHANNEL NUMBER to [MEM]. 1.004 DISP Set the MEASUREMENT RANGE to 00. 1.004 DISP Set the RECORDING RANGE to -20.00mV to 20.00mV. 1.004 DISP 1.004 DISP Refer to instruction manual if unfamiliar w/keyboard 1.005 DISP Using the keyboard on the side of the UUT.. 1.005 DISP 1.005 DISP Set the DISPLAY to MAN DATA. 1.005 DISP 1.005 DISP USING the < > keys SELECT CHANNEL [MEM]. 1.006 DISP Connect the 5520 to the UUT Channel [MEM] as follows: 1.006 DISP [32] 5520A NORMAL HI to UUT A - TOP INPUT TERMINAL 1.006 DISP [32] 5520A NORMAL LO to UUT B - BOTTOM INPUT TERMINAL 1.007 5520 -20.00mV S 2W 1.008 DISP Adjust the channel [MEM] ZERO adjustment for a 1.008 DISP recording at the LEFT EDGE of the chart (0). 1.008 DISP 1.008 DISP NOTE: The 100% point will be moved the same amount. 1.009 5520 20.00mV S 2W 1.010 DISP Adjust the channel [MEM] SPAN adjustment for a 1.010 DISP recording at the RIGHT EDGE of the chart (100). 1.010 DISP 1.010 DISP NOTE: The zero point is slightly affected by span adj. 1.011 5520 -20.00mV S 2W 1.012 OPBR Is the ZERO at the LEFT EDGE of the chart +/- 1div? 1.013 JMPF 1.007 1.014 5520 20.00mV S 2W 1.015 OPBR Is the 100% at the RIGHT EDGE of the chart +/-1 div? 1.016 JMPF 1.009 1.017 5520 * S 1.018 END